In-vitro investigation of primary stability of orthodontic mini implants with different lengths using resonance frequency analysis


Original Article

Author Details : Amruta Mantri Chandak*, Suchita Daokar Tarvade, Mitali Sharma, Harsha Jairaj Kaurani

Volume : 6, Issue : 4, Year : 2022

Article Page : 146-151

https://doi.org/10.18231/j.jco.2022.029



Suggest article by email

Abstract

Introduction: Out of various ways of gaining anchorage, mini implants are gaining attention due to its minimum compliance and maximum curative effects.Mini implants are considered successful if they have sufficient primary stability that comes from mechanical interlocking of mini implant with the bone. Factors which influence the implant stability are bone physiology, implant size, shape and surface characteristics. Considering all other factors within anatomic constraints, what may play a role in primary stability of mini implants is length. Literature search to correlate mini implant length with primary stability was found to be insufficient and hence this study was undertaken.
Materials and Methods: In this study, two goat jaws were subjected to spiral 3D CT scan and the areas with D2 bone density were identified and marked. In these D2 density marked areas, 30 implants of 1.5 x 6mm (GROUP A) and 30 implants of 1.5 x8mm (GROUP B) were placed. Their stability was measured by Radio Frequency Analysis using Osstell ISQ device with its Smart Peg. A connector was fabricated to make the fit of smart peg compatible with the head of mini implant. Readings were made in 5 different directions for each implant and their average value was considered as final reading.
Results: It was noted that primary mini implant stability is significantly higher (p 0.034) with GROUP B implants than with GROUP A implants.
Conclusion: The increased length of mini-implants positively affects its primary stability and should be taken into consideration when implants are used as anchorage devices.


Keywords: Resonance Frequency Analysis, Primary stability, Length, D2 density


How to cite : Chandak A M, Tarvade S D, Sharma M, Kaurani H J, In-vitro investigation of primary stability of orthodontic mini implants with different lengths using resonance frequency analysis. J Contemp Orthod 2022;6(4):146-151


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International, which allows others to remix, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.







Article History

Received : 20-06-2022

Accepted : 01-11-2022


View Article

PDF File   Full Text Article


Downlaod

PDF File   XML File   ePub File


Digital Object Identifier (DOI)

Article DOI

https://doi.org/10.18231/j.jco.2022.029


Article Metrics






Article Access statistics

Viewed: 785

PDF Downloaded: 221