Introduction
Anchorage, defined as a resistance to unwanted tooth movement, is a prerequisite for the orthodontic treatment of dental and skeletal malocclusions.1 The growing demand for orthodontic treatment methods that require minimal compliance and provide maximal anchorage control, particularly for adults, has led to the expansion of implant technology in orthodontics.2
The basic requirement for the success of orthodontic mini-implants is sufficient primary stability. Primary stability basically comes from mechanical interlocking with the cortical bone when the mini-implant is placed. It is influenced by bone quality and quantity, surgical technique, and screw geometry.3
Even though mini-implants are being extensively used, a major drawback of Orthodontic mini-implants is their failure rate. The reported failure rate of mini-implants varies from 6.6 to 16.1%.4
Mini-implant placement generates stresses and strains along the length of the screw that damage the surrounding bone. Too much damage can lead to micromotion of the implant and early loss due to lack of stability.5
Research to improve the stability of mini-implants are on rise.
Platelet rich fibrin (PRF) is new generation platelet concentrate, a healing biomaterial, that has shown great potential for bone and soft tissue regeneration, without inflammatory reactions and may be used alone or in combination with bone grafts, promoting haemostasis, bone growth, and maturation. It consists of an autologous leukocyte-platelet-rich fibrin matrix6 composed of a tetra molecular structure, with cytokines, platelets, and stem cells within it,7, 8 which acts as a biodegradable scaffold that favours the development of micro-vascularization and is able to guide epithelial cell migration to its surface.7, 8, 9, 10
PRF may enhance the stability of mini-implants due to its healing property. Therefore, this study was undertaken to evaluate and assess the efficacy of PRF on the stability of orthodontic implant.
The present study investigated the effect of PRF on stability of mini-implants. A randomized clinical control trial was performed, with a focus on measuring the stability of mini-implant within the maxilla without loading following the extraction of first premolar teeth, to assess the effect of PRF on the miniimplant stability after 8 weeks using resonance frequency analysis, in comparison to that of the randomly allocated control group. The secondary outcomes expected was to evaluate the optimal loading time after the day of insertion of orthodontic micro-implants.
Materials and Methods
Trial design
The study was a single-centered, split-mouth randomized control trial to investigate the effectiveness of PRF on the stability of orthodontic mini-implant using resonance frequency analysis and secondary outcome.
Participants, eligibility criteria, and settings
The sample comprised patients who were undergoing comprehensive fixed orthodontic treatment with premolar extraction. Subjects who fulfilled the inclusion criteria (Table 1) were randomly allocated to 1 of the 2 groups. The lottery method of the simple random sampling technique was followed after informed consent was obtained from all the study subjects. On the one side was the test group (PRF group), and on the other side was the positive control (Non PRF) group. A total of 16 subjects were included in the study, and overall, 32 mini-implants were undertaken for assessment (16 mini-implants coated with PRF and 16 without PRF coating).
Sample size
A pilot study was done to determine the sample size of the 2 different groups of the study.11 The values obtained from each group were calculated and the effect size was obtained using G* power 3.1.9.2. Subsequently, we obtained a sample size of 13 per group. To avoid unexpected errors, outcomes and dropout, the size was rounded up to 16 per group. The samples were split into 2 groups in which interventions were carried out (Figure 1):
Group 1: Among the patients in this group, the implants were coated with the PRF before insertion.
Group 2: Among the patients in this group, were not coated with PRF before insertion.
The fresh PRF was made for each patient by following the given procedure:
Around 2 ml of whole venous blood was drawn from cubital vein after torniquet application in upper arm (Figure 1). A 5ml syringe with 26-gauge needle was used for blood extraction and blood was collected in each of the two sterile vacutainer tubes (test tube) of 5ml capacity without anticoagulant. Following blood collection, the vacutainer tubes were then placed in a centrifugal machine (Figure 2) at 3000 revolutions per minute (rpm) for 10 minutes. The resultant product (Figure 3) consisted of following layers:
Top layer consists of straw-coloured cellular plasma.
Middle layer consists of platelet rich fibrin.
Bottom layer consists of R.B.C.
P.R.F and red corpuscle base was extracted using a sterile forceps from straw-coloured cellular plasma and then with help of a sterile scissor P.R.F. was cut-off from R.B.C and transferred onto a sterile dappen dish.16 implants were coated with PRF using zero size brush (Figure 4).
All the 18 titanium- Grade 5(Ti-6AI-4V) implants (1.5 x 6mm) were then inserted 8 mm from the alveolar crest between the upper first molar and premolar with a self-drilling method at 90º
The ISQ was measured using resonance frequency analyser at 6 different time intervals – at time of insertion(T0), 24 hours (T1), 2 weeks (T2), 4 weeks (T3), 6 weeks(T4) and 8 weeks(T5) after insertion in two directions i.e., mesiodistal direction (D1) and occlusogingival direction (D2). (Figure 4, Figure 5)
Result
Statistical analysis
The observations obtained in the study were subjected to statistical analysis, so as to get their interpretation. Data was coded, transferred and analysed on SPSS version 19. Mann Whitney test was used to compare the mean ISQ values measured in the two different directions and the ANOVA test was used for intragroup and intergroup comparison. Level of significance was set at p<0.05.
Table 1
Table 2
Table 3
Table 4
Table 5
Table 6
Statistically insignificant findings were observed when direction D1 i.e., Mesio-distal direction when compared to direction D2 i.e., Occluso-gingival direction using Mann Whitney test at T0, T1, T2, T3, T4 and T5 in both the groups. (Table 2, Table 3)
Statistically significant findings were found when group A (experimental group) was compared to group B (control group) using ANOVA test (p<0.05). The stability of implant of group A at each time interval was greater than group B. (Table 4, Table 5) (Figure 6)
The difference in mean ISQ for the intragroup comparison by using ANOVA test was also found statistically significant. The implant stability was found to be high at T0 and reduces till T2 and then increases through T3 to T5 in both the groups. The stability was highest at T5. (Table 6, Table 5)
Discussion
In this study, statistically significant findings were found when group A (Experimental group) was compared to group B (Control group) using ANOVA test (p<0.05).
The stability of implant of group A at each time interval was greater than group B. This shows that implants coated with PRF have increased its stability suggesting PRF can significantly enhance the implant stability. This may be because it consists of an autologous leukocyte-platelet-rich fibrin matrix6 composed of a tetra molecular structure, with cytokines, platelets, and stem cells within it,7, 8 which acts as a biodegradable scaffold9 that favours the angiogenesis and is able to guide epithelial cell migration to its surface.7, 10 The mechanism which was followed here was that, fibrinogen which was initially concentrated in the high part of the tube, combined with the circulating thrombin due to centrifugation, to form fibrin. A fibrin clot was then obtained in the middle of the tube, just between the red corpuscles at the bottom and acellular plasma at top. Platelets were trapped massively in the fibrin meshes.5
In intragroup comparison of ISQ values at different time intervals were analyzed using ANOVA Test it was observed that the implant stability was higher at T0 and reduces till T2 and increasing through T3 to T5. The stability was lowest at T2 and highest at T5.
This shows that the primary stability of implant is best at the time of insertion, however the stability gradually declines from 24 hours till 2nd week after implant insertion which may be due to inflammatory response of the adjacent tissues to the trauma caused at the time of implant placement. The stability then starts increasing after 2nd week of implant stability shows that healing sets in after the initial inflammatory response and improves the stability. The highest ISQ value was shown at 8 weeks of duration. Suggesting that the best time to load the implants will be immediately after the placement of the implant or after 8th week of insertion of implant.
With above findings we can also conclude that the best time of loading the implant is at the time of insertion i.e., immediate loading is recommended.
The result of this study could not be compared with any other studies as there were no studies found regarding use of PRF with mini-implants in the literature.
Conclusion
In this study, PRF coating has shown a promising result in terms of increase in stability of implants. So, with the use of PRF, the stability of mini-implants can be increased by chairside with minimally invasive procedure. There is a scope to evaluate the effect of PRF of micro-implants on primary stability with loading the mini-implants with an increased sample size for long period of time.
The limitation of this study is that it includes the limited sample size and the primary stability was measured without loading the micro-implants which would also affect the primary stability. Another drawback is that only a single type of micro-implants from a same company was used. Further investigations using different types of mini-implants at different positions is recommended.